Summary
Type: SPE
Application #: 304180
Catalog#: 730075 - SPE Cartridges, Chromabond, Silica, 6ml, PP, 1000mg, 30/pk
Detection: HPLC in reference to Appl. No: 121661 and 121662
Determination limit: 0.05 mg/kg
Conditions: Column conditioning: 2 ml n-pentane (Do not let run the column dry.)
Sample aspiration: sample solution is passed through the column by vacuum or pressure (appr. 3 ml/min)
Washing: 20 ml n-pentane, drying for 5 min by vacuum air stream
Elution: 2 ml methanol / water, 20:80, v/v
Authors: Heisler A et al
Source: GIT Labor-Fachzeitschrift 5/2004 504-505